Intermittent regurgitation caused by incomplete leaflet closure of the Medtronic ADVANTAGE bileaflet heart valve: Analysis of the underlying mechanism

German Heart Center Munich, Department of Cardiovascular Surgery, Munich, Germany

Received 7 April 2009; received in revised form 28 August 2009; accepted 1 November 2009. published online 01 February 2010.

Objective

Clinical echocardiographic assessments of the Medtronic ADVANTAGE (Medtronic Inc, Minneapolis, Minn) prosthesis in the aortic position revealed a phenomenon identified as “intermittent regurgitation.” An in vitro investigation was initiated to identify the underlying mechanism.

Methods

In a pulse duplicator environment, 6 ADVANTAGE size 23 aortic valves were analyzed. Leaflet motion and flow through the valves were documented using echocardiography with color Doppler flow, digital high speed imaging, and flow meter assessment.

Results

Intermittent regurgitation could be reproduced in all 6 of the tested valves within limited ranges of flow, pressure, and valve orientation. By virtue of high-speed imaging, the mechanism underlying intermittent regurgitation was identified. During intermittent regurgitation, the leading edge of the second-to-close leaflet makes contact with the chamfer on the leading edge of the first-to-close leaflet. The fluid closing forces working on the first-to-close leaflet prevent it from shifting back so that the leading edge of the second-to-close leaflet remains positioned against the chamfer of the first-to-close leaflet. In this position, the major radius of the second-to-close leaflet does not reach the housing's major radius. Therefore, a crescent-shaped gap remains between the leaflet tip of the second-to-close leaflet and the housing major radius during all or part of diastole. The regurgitant fraction can increase from a normal range of 6% to 25% during an intermittent regurgitation beat.

Conclusions

In vitro intermittent regurgitation can be induced in the size 23 aortic ADVANTAGE valve under a limited range of conditions. To avoid possible misinterpretations, the phenomenon must be known in detail by all physicians dealing with patients with an ADVANTAGE valve.

CTSNet classification: 35.2, 35.3.4, 36.2

Abbreviation and Acronym: IR, intermittent regurgitation

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Funding: In vitro measurements were performed by Medtronic Inc, Minneapolis, Minnesota.

 Disclosures: Walter Echinger reports consulting/lecture fees and grant support from Medtronic and St. Jude. Andrea Hutter reports lecture fees and grant support from Edwards. Rudiger Lange reports consulting fees and grant support from Medtronic and Edwards.

 The German Heart Center is part of the investigating institutions in the ADVANTAGE trial initiated by Medtronic Inc.

PII: S0022-5223(09)01427-5

doi:10.1016/j.jtcvs.2009.11.001

The Journal of Thoracic and Cardiovascular Surgery
Volume 140, Issue 3 , Pages 611-616, September 2010