The Journal of Thoracic and Cardiovascular Surgery
Volume 139, Issue 6 , Pages 1494-1500, June 2010

Fifteen years of experience with ATS mechanical heart valve prostheses

  • Akira Sezai, MD

      Affiliations

    • Department of Cardiovascular Surgery, Nihon University School of Medicine, Tokyo, Japan
    • Corresponding Author InformationAddress for reprint: Akira Sezai, MD, the Department of Cardiovascular Surgery, Nihon University School of Medicine, 30-1 Oyaguchi-kamimachi Itabashi-ku, Tokyo, 173-8610, Japan.
  • ,
  • Mitsumasa Hata, MD

      Affiliations

    • Department of Cardiovascular Surgery, Nihon University School of Medicine, Tokyo, Japan
  • ,
  • Tetsuya Niino, MD

      Affiliations

    • Department of Cardiovascular Surgery, Nihon University School of Medicine, Tokyo, Japan
  • ,
  • Isamu Yoshitake, MD

      Affiliations

    • Department of Cardiovascular Surgery, Nihon University School of Medicine, Tokyo, Japan
  • ,
  • Yuji Kasamaki, MD

      Affiliations

    • Department of Cardiology, Nihon University School of Medicine, Tokyo, Japan
  • ,
  • Atsushi Hirayama, MD

      Affiliations

    • Department of Cardiology, Nihon University School of Medicine, Tokyo, Japan
  • ,
  • Kazutomo Minami, MD

      Affiliations

    • Department of Cardiovascular Surgery, Nihon University School of Medicine, Tokyo, Japan

Received 14 April 2009; received in revised form 17 June 2009; accepted 16 July 2009. published online 11 September 2009.

Background

ATS Medical, Inc, developed a mechanical heart valve that has been in use since 1992. In this article, we present the results of 15 years of follow-up of patients who have undergone ATS heart valve replacement at our hospital.

Methods and Results

We performed ATS heart valve replacements on 231 patients between September 1993 and March 2008. Our operative mortality rate for the study period was 2.2%. The survival for postoperative thromboembolic events was 0.29%/pt-y for aortic valve replacement, 0.48%/pt-y for mitral, 0.80%/pt-y for double valve replacement, and overall 0.44%/pt-y. The survival after bleeding events was 0.29%/pt-y for aortic valve replacement, 0.16%/pt-y for mitral, 0%/pt-y for double valve replacement, and overall 0.19%/pt-y. Patient–prosthesis mismatch, as determined by echocardiography, was found in 83.3% of patients at 19 mm, but other sizes showed good valve function. Prosthetic valve noise was undetectable in 92.8% of patients, and quality of life was excellent.

Conclusions

Few prosthetic valve–related complications were seen with ATS heart valve replacements in this study, and the follow-up results were favorable. The international normalized ratio was maintained in the range 1.6 to 2.0 in patients with aortic valve replacement in sinus rhythm. Not only bleeding events, seen at a rate of 0.19%/pt-y, but also thromboembolic events, at 0.44%/pt-y, were low when compared with conventional mechanical valves. Prosthetic valve noise is low, and this appears to be an excellent mechanical valve from the quality of life standpoint. The ATS valve has an excellent safety profile when compared with other mechanical valves.

CTSNet classification: 35, 35.3

Abbreviations and Acronyms: AP, advanced performance, AVR, aortic valve replacement, DVR, double valve replacement (AVR + MVR), LDH, lactic acid dehydrogenase, MVR, mitral valve replacement, PPM, prosthesis–patient mismatch, PVL, paravalvular leakage, QOL, quality of life, SJM, St Jude Medical

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 This study was supported by an Osawa Research grant from Nihon University.

 Disclosures: None.