Hemodynamic effects of partial ventricular support in chronic heart failure: Results of simulation validated with in vivo data

  • Deborah Morley, PhD

      Affiliations

    • Department of Clinical Affairs, CircuLite, Inc, Hackensack, NJ
    • Corresponding Author InformationAddress for reprints: Deborah Morley, PhD, CircuLite, Inc, Clinical and Regulatory, 401 Hackensack Ave, Hackensack, NJ 07601
  • ,
  • Kenneth Litwak, DVM, PhD

      Affiliations

    • Department of Surgery, University of Louisville, Louisville, Kentucky
  • ,
  • Paul Ferber, BS

      Affiliations

    • Center for New Media Teaching and Learning, Columbia University, New York, NY
  • ,
  • Paul Spence, MD

      Affiliations

    • Louisville, Ky
  • ,
  • Robert Dowling, MD

      Affiliations

    • Department of Surgery, University of Louisville, Louisville, Kentucky
  • ,
  • Bart Meyns, MD, PhD

      Affiliations

    • Department of Surgery, University of Leuven, Leuven, Belgium
  • ,
  • Bartley Griffith, MD

      Affiliations

    • Department of Surgery, University of Maryland, Baltimore, Maryland
  • ,
  • Daniel Burkhoff, MD, PhD

      Affiliations

    • Center for New Media Teaching and Learning, Columbia University, New York, NY
    • Cardiovascular Research Foundation, Orangeburg, NY

Received 22 November 2005; received in revised form 9 June 2006; accepted 7 July 2006.

Objective

Current left ventricular assist devices are designed to provide full hemodynamic support for patients with end-stage failing hearts, but their use has been limited by operative risks, low reliability, and device-related morbidity. Such concerns have resulted in minimum use of left ventricular assist devices for destination therapy. We hypothesize that partial circulatory support, which could be achieved with small pumps implanted with less-invasive procedures, might expand the role of circulatory support devices for treatment of heart failure.

Methods

We examine the hemodynamic effects of partial left ventricular support using a previously described computational model of the cardiovascular system. Results from simulations were validated by comparison with an in vivo hemodynamic study.

Results

Simulations demonstrated that partial support (2-3 L/min) increased total cardiac output (left ventricular assist device output plus native heart output) by more than 1 L/min and decreased left ventricular end-diastolic pressure by 7 to 10 mm Hg with moderate-to-severe heart failure. Analyses showed that the hemodynamic benefits of increased cardiac output and decreased left ventricular end-diastolic pressure are greater in less-dilated and less-dysfunctional hearts. Both the relationships between ventricular assist device flow and cardiac output and ventricular assist device flow and left atrial pressure predicted by the model closely approximated the same relationships obtained during hemodynamic study in a bovine heart failure model.

Conclusions

Results suggest that a pump with a flow rate of 2 to 3 L/min could meaningfully affect cardiac output and blood pressure in patients with advanced compensated heart failure. The development of small devices capable of high reliability and minimal complications that can be implanted with less-invasive techniques is supported by these findings.

CTSNet classification: 17, 22, 27

Abbreviations and Acronyms: CO, cardiac output, CRT, cardiac resynchronization therapy, EDP, end-diastolic pressure, LA, left atrium, LAP, left atrial pressure, LVAD, left ventricular assist device, LVEDP, left ventricular end-diastolic pressure, VAD, ventricular assist device

 

 The study was supported by a grant from CircuLite, Inc. Deborah Morley reports CircuLite employment and stock options. Kenneth Litwak reports CircuLite consulting fees. Paul Ferber Reports Atria Medical, CircuLite, Dupont, and Transoma consulting fees. Paul Spence reports CircuLite consulting fees, equity, and patents. Robert Dowling reports CircuLite consulting fees. Bart Meyns reports Abiomed lecture fees, CircuLite consulting fees and equity, and grant support from Abiomed and CircuLite. Bartley Griffith reports WorldHeart consultant fees. Daniel Burkhoff reports Accelerated Technologies and Impulse Dynamics consulting fees; Abiomed, CircuLite, and Impulse Dynamics equity; and grant support from CircuLite. All of the companies listed have direct or indirect commercial interests in implantable cardiac assist devices.